3-in-1 microvolume spectrophotometer Genova Nano - JENWAY

Sku: 723475
Jenway
The strong points
  • Ideal for DNA, RNA, and protein measurements Spectral scanning across the full wavelength range from 198 to 1000nm

  • Detection of DNA concentration up to 2ng/µL

  • Easy and quick to clean

€8,044.49 €9,653.39
- +
On order

Microvolume spectrophotometer, for life sciences and standard applications in one instrument.

This microvolume spectrophotometer is an innovative spectrophotometer for the sciences thanks to Jenway's microvolume measurement accessory.
The Genova Nano microvolume spectrophotometer measures small sample volumes, up to 0.5 µL with unparalleled precision, reproducibility, and speed. It has the ability to measure small sample volumes, reducing the need for dilution. This measurement mode allows for the preservation of valuable samples without the use of cuvettes. By wiping a microfiber cloth over the reading head, any sample traces are eliminated, reducing the time between measuring two samples and consequently increasing the analysis throughput.

Genova Nano single beam micro-volume spectrophotometer, UV/visible (198-1000 nm) delivered with micro-volume accessory, universal power cable, 4GB USB key, calibration certificate, and user manual.

Résolution Non connu
Type de spectrophotomètres UV / Visible
Livré avec logiciel Non
Bande passante 5 nm
Source Lampe Xénon
Gamme spectrale 198 - 1000 nm
Faisceau Mono
Sorties USB
Caractéristiques techniques
Wavelength range: 198 to 1000 nm
Wavelength accuracy: ± 2nm
Bandwidth: 5 nm
Optical path: 0.2 or 0.5 mm (manual or automatic selection)
Absorbance range: -15 to 125 A (equivalent to 10 mm)
Absorbance accuracy: ± 2% at 260 nm
Absorbance precision: < 0.005 A between 0 and 1 A (at 260 nm and 0.5 mm)
Maximum concentration: 6,000 ng/µl (dsDNA) (at 0.2 mm)
Limit of detection: 2.0 ng/µl (dsDNA) (at 0.5 mm)
Measurement time: < 6.5 seconds
Minimal sample volume: 0.5 µl (at 0.2 mm), 1.0 µl (at 0.5 mm)
Maximum sample volume: 5.0µl
DNA measurement modes: dsDNA, ssDNA, RNA, Oligonucleotides, 260/280, 260/230, variable ratio
Protein measurement modes: Bicinchoninic acid assay (BCA), Bradford method, Lowry method, biuret method, direct UV
Reading head: Chemically inert quartz lens and stainless steel
Light source: Xenon lamp with "Press To Read" (PTR) technology
Dimensions (l x w x h): 275 x 400 x 220 mm
Weight: 7.7 kg