Vacuum oven VO - MEMMERT
SKU | Capacité | Price | Available | Amount Basket | ||
---|---|---|---|---|---|---|
707443 | 29 L |
€8,049.29
€9,659.15
|
On order |
|
||
707444 | 49 L |
€9,088.79
€10,906.55
|
On order |
|
Pressure
Vacuum range: 5 to 1100 mbar
Pressure regulation: Digital electronic regulation of the vacuum pump pressure with speed control. Vacuum, air, and inert gas pipes are made of material 1.4571 (ASTM 316 Ti). Programmable air inlet with digital control.
Admissible final vacuum: 0.01 mbar
Maximum leakage rate: 0.01 bar/h
Pump control: Speed regulation, optimized flushing procedures for pump membranes, and signal output for pump ON/OFF.
Connection: Vacuum connection with small DN16 flange, and gas inlet with small DN16 flange
Temperature
Temperature: Temperature measurement from a Pt100 probe in 4-wire technology, separate for each heating plate
Useful temperature range: At least 5 °C above ambient temperature up to +200 °C
Display resolution for setpoint and actual temperature values: 0.1 °C
Control Technique
Adjustable parameters: Temperature (Celsius or Fahrenheit), vacuum, program duration, time zones, daylight saving time
Language configuration: German, English, French, Spanish, Polish, Czech, Hungarian, Italian
ControlCOCKPIT: TwinDISPLAY. Digital PID controller with adaptive multifunction microprocessor with 2 high-resolution TFT color displays
Clock: Digital timer with stop time programming and duration setting from 1 minute to 99 days
SetpointWAIT function: Process starts only when set temperature is reached
Communication
Interface: Ethernet LAN, USB
State recording: Program cycle recording in case of power failure
Programming: Programming, management, and program download from Ethernet interface or USB port using AtmoCONTROL software
Safety
Thermal safety device:
Mechanical temperature limiter TB with protection class 1 (DIN 12880) for heating cutoff at about 10 °C above the maximum device temperature
AutoSAFETY:
Integrated additional thermal safety device with alarm for automatic monitoring of setpoint, which automatically monitors the set value within a freely defined range, alarms for exceeding upper/lower thresholds, and cuts off heating in case of overheating or cooling unit in case of undercooling.
Automatic diagnostic system: For detecting temperature control anomalies
Alarm: Visual and audible
Heating concept
Direct VO heating: MLC regulation (multi-level control) by microprocessor with fuzzy logic emulation and volumetric optimization (performance adaptation based on the loading element volume), with local thermal probes and separate operation for each heating plate.
Heating plate: 2 connections for heating plate at the back wall (first and second level)
Alimentation | 230 V |
Plage de régulation de la température | 5 to 90 °C |
Stabilité | ±0,3 °C |
Type d'affichage | Ecran numérique |
Caractéristiques techniques
Basic Equipment
Factory Calibration Certificate: for a temperature of +160 °C at a pressure of 20 mbar for each thermo-plate delivered with the oven. Drawers: 1 aluminum thermo-plate 3.3547 (ASTM B209) with heating for large surfaces Inner casing made of stainless steel Material: Hermetic and sealed welding, made of stainless steel extremely resistant to corrosion, material number 1.4404 Interior: Additional interior coating in stainless steel 1.4404 (removable for cleaning), side walls with support rails and anti-tilt device, coating of the upper interior wall for protection against turbulence caused by ventilation Volume: 29 l Dimensions: L(A) x H(B) x D(C): 385 x 305 x 250 mm Max. number of shelves: 2 Maximum device load: 40 kg Max. load per shelf: 20 kg Outer casing made of structured stainless steel Dimensions: L(D) x H(E) x D(F): 550 x 607 x 400 mm (D +38mm excluding handle) Door: Transparent glass door, interior with 15 mm thick safety glass mounted on a spring, exterior with anti-shatter protection Outer casing: Back wall made of fully galvanized steel sheet Electrical Data Voltage: Power (only valid for max. equipment) at 230 V, 50/60 Hz Environmental Conditions Ambient Temperature: +5 ºC to +40 ºC Installation: A minimum gap of 15 cm should be provided between the wall and the rear panel of the enclosure. The gap should not be less than 20 cm from the ceiling and 8 cm on the sides of the device. Relative Humidity: max. 80%, non-condensing Overvoltage Category: II Pollution Degree: 2 |