Heating element for QWS4 distiller
Heating element for Pyrex Quickfit® QWS4 water distiller (Jeulin reference 707 231). REMINDER: After a certain period of use, scale may build up on the heating element and inside the boiler. To ensure the proper functioning of the device, the scale must be regularly removed. The time interval between 2 cleanings depends on the hardness of the feed water. In regions where the water is very hard, it may be necessary to clean the distiller every week. In regions where the water is softer, several weeks can pass between 2 cleaning operations. Significant scaling will reduce the quality of the distilled water and the lifespan of the heating element. The appearance of white deposits on the heating element and/or the boiler is a sign that descaling is necessary. Scale dissolves perfectly in a 10% diluted formic acid solution. Some household descaling solutions can also be used. Item supplied without electrical plug. Connection to the power supply to be done by the user. 1 year or 1000 hours warranty.
Caractéristiques techniques
Item provided without electrical plug.
Connection to power supply to be done by you. 1 year or 1000 hours warranty. |