FiveEasy TM F30 Benchtop Conductivity Meter - METTLER TOLEDO

Sku: Pr_259434
Mettler Toledo
The strong points
  • Intuitive solution for conductivity measurement in the laboratory

  • Measurements made easy

  • Compact and clever design

From €1,030.00 €1,236.00 €0.00 eco-part
Configuration

Model
SKU Modèle Price Available Amount Basket
702276 F30-Meter
€1,030.00 €1,236.00
On order
+
702277 F30-Standard
€1,200.00 €1,440.00
On order
+

Combining compact design and intuitive operation, the new FiveEasy conductivity meters ensure quality measurements at the best price in your laboratory! Intuitive operation - Simplified measurement Equipped with a large well-arranged screen for easy result reading, these instruments provide ultra-fast measurements thanks to intuitive menus and buttons. Compact design - Minimal footprint Thanks to their compact design, the instruments in the Five series take up very little space on the bench. The electrode holder can be easily removed and attached to the side of the instrument after use. Large screen - Simultaneous display of information Well-arranged, the screen shows explicit icons indicating all the required information, such as measurement data, electrode status, and current calibration.

Ref: - F30-Meter Ideal for: Applications requiring a specific sensor not included in the kit Includes the Conductimeter delivered with user manual, quick guide, and power adapter Ref: - F30-Standard Ideal for: Aqueous samples with medium to high conductivity Includes the device and accessories as for the FP30-Meter with an additional LE703 conductivity cell, electrode holder, 2 x 1413 μS/cm calibration sachets, and 2 x 12.88 mS/cm calibration sachets

Gamme de température 0 to 100 °C
Mesure de pH Précision Non connu
Résolution Plage automatique
Étalonnage Automatique
Gamme de mesure 0,01 µS/cm à 200 mS/cm
Type Paillasse
Caractéristiques techniques
Conductivity
Measurement range: -0.01 µS ... 200 mS
Resolution: automatic range
Error limit: ± 0.5% of the measured value

TDS
Measurement range: -0.01 mg/L ... 200 g/L
Resolution: automatic range
Error limit: ± 0.5% of the measured value

Temperature
Measurement range: -0 ... 100 °C
Resolution: 0.1 °C
Error limit: ± 0.5 °C

Temperature Compensation: linear from 0.00 %/°C to 10.00 %/°C, reference temperature: 20 and 25 °C
Calibration: 1 point, 3 predefined standards
Data Storage: current calibration
Connection: mini-DIN (electrode)
Power Supply: 12 V power adapter (included)