BE filtration tank (for acidic vapors)
To be connected using a flexible hose (provided) to the ventilation outlet of the cabinet. The unit is placed on top of it. The treated air is reintegrated into the ambient air.
Filter
molecular filter based on high-performance activated carbon ensuring filtration compliant with standard NF X 15-211:
- A filtration efficiency standard ensuring the protection of people's health.
- Its objective: to only reject 1% of what is permissible, which is 100 times below the official threshold determined by INRS. This efficiency is not excessive when one is daily exposed in a laboratory to an atmosphere containing chemical pollutants (gases, vapors, aerosols, etc.).
Regular replacement (once a year) is recommended to ensure this efficiency and the proper functioning of your filtration equipment.
Équipement(s) | Caisson à filtration |
Caractéristiques techniques
Steel pump. Motor with turbine made of anticorrosive composite material. 230 volts single phase. Double fuse base, maximum flow rate: 74 m3/h, power 20 W, sound level of 49 dB(A). Supplied with MF/BE filter. Dimensions (H x W x D): 300 x 410 x 430 mm (+ 38 mm flange). French manufacturing.
|