Display: Alphanumeric 4 lines
Precision: ± 1 nm - Repeatability: ± 0.1 nm
Spectrum acquisition: Automatic
Measurement modes: Absorbance, Transmission, Concentration, With embedded software: kinetics, calibration range (10), method storage (30).
Photometric range: Absorbance: -0.2 to 3.5 A / Transmission: 0 to 200%
Photometric precision: ± 0.005 to 1 A nm (delivered with calibration certificate)
Light source: Long-life Tungsten (10-year warranty *)
Monochromator: Ebert type (sealed)
Detector: Photodiode (sample illuminated with monochromatic light, avoiding heating, drift, and evaporation as in a double-beam.
Standard: Holmium oxide standard filter integrated, with automatic validation utilities.
Universal cuvette holder for:
- Cuvettes with 5 to 50 mm optical path length
- Tube with 13 to 25 mm diameter
Outputs:
- USB as standard
- Parallel port for printer is included
- Analog 0 to 1 Volt (optional)
- RS232 serial port (available on request)
Software: optional
Power supply: 115/230 V - 50/60 Hz
Dimensions: 400 x 300 x 145 mm
Weight: Approximately 10 kg (depending on options)Sensitivity: up to 5.10-8 g/l of quinine - Detector: High sensitivity photodiode (10-year warranty for the detector which can be exposed to full light).
Cuvettes: "Standard" cuvettes for 10 mm optical path fluorescence. Measurements can be performed on fluorescence cuvettes (polished / transparent side faces) made of quartz, glass, or transparent plastic up to 350 nm. Any surface defects on cuvette side faces not intended for fluorescence may be acceptable, however, they reduce measurement sensitivity, and cuvettes must always be used in the same position.
Connection: USB 2 - Computer with Windows 10 or 11 required for operation (not provided). Free software updates for the entire instrument's lifespan (Sustainable Spectroscopy concept).