conductivity meter sensION + EC7

Sku: 610054
Hach Lange
The strong points
  • An "All in One" device One-piece unit including electrode holder, stirrer, solutions and calibration beakers Simple, reliable and compact Delivered with a conductivity cell Compatible with Radiometer Analytical sensors

€1,261.00 €1,513.20

- Magnetic agitator
- Standard solutions: 147 and 1413 µS/cm and 12.88 mS/cm
- Support for cell
- Vials for calibration with stirring magnet
- 5070 conductivity cell (for general use)
- Manual

Mesure de pH Précision Non connu
Résolution 0,01 µS à 1 mS/cm
Étalonnage Manuel
Gamme de mesure 0,01 µS/cm à 500 mS/cm
Type Paillasse
Sorties Non communiqué
Caractéristiques techniques
Settings :
Conductivity: 0.01/0.001*µS/cm at 500/100 **ms//cm* with Cte=0.1 cm-1 with Cte=10 cm-1.
Resistivity : 0 to 100 MOhms.
Salinity: 5.85 mg/l at 311.1 g/l NaCl.
Temperature: -20 to 150°C (-4 to 302°F).
Measurement error (± 1 digit). Repeatability (± 1 digit).
Conductivity =0.5%.
Conductivity ± 0.1%.
Resistivity =0.5%.
Resistivity ± 0.1%.
Salinity =0.5%.
Salinity ± 0.1%.
Temperature =0.2°C (0.4°F).
Temperature ± 0.1°C (0.1°F).
Temperature compensation:
Automatic with a Pt 1000 or 10 kOhms temperature probe.
Manual if no T° sensor connected.

Conductivity calibration:
1, 2 or 3 calibration points: Choice of standards in the range 147 µS/cm, 1413 µS/cm, 12.88 mS/cm and 111.8 mS/cm (at 25°C).
Manual entry of the cell constant. Possible return to a default calibration. Programmable calibration reminder from 0 to 99 days.

Conductivity temperature correction:
Linear correction factor from 0 to 9.99%/°C (2% by default).
Reference temperature 20 or 25°C.

Measurement mode: Stable or continuous.

Display: Graphic, liquid crystal and backlit. Multilingual text menu: FR, GB, ES, DE and PT.

Connection: 8-contact telephone socket for conductivity cell with integrated T° sensor. RCA socket for stirrer.

Housing characteristics: Waterproof housing (IP42). Dimensions 350 x 200 x 110 mm / Weight 1.1 kg.